松江分析儀

來源: 發(fā)布時(shí)間:2023-02-25

氧氮?dú)浞治鰞x的常見故障及解決方法供電正常、通訊正常,點(diǎn)擊確認(rèn)鍵后分析儀不工作。這是沒有水流,爐子溫度太高或儀器通道電壓不正常。如沒有水流,爐子溫度太高這些信息會(huì)顯示在顯示器畫面上,但沒有信息顯示說明這兩項(xiàng)正常。接下來檢查儀器通道零位電壓,如果比±3V高出1V以上,可能是因?yàn)闅馄靠樟?,或者是空氣進(jìn)入到分析儀中。檢查并更換化學(xué)試劑,如果有空氣進(jìn)入熱導(dǎo)池里,熱導(dǎo)池的電壓就會(huì)<-6V,此時(shí)打開右面的門,堵住爐子氣體進(jìn)口,10s后,熱導(dǎo)池電壓值必然升高。經(jīng)過逐一排查,**終確認(rèn)堿石棉有問題,更換后儀器正常。上海禹重實(shí)業(yè)有限公司為您提供分析儀,歡迎新老客戶來電!松江分析儀

松江分析儀,分析儀

Procedure-SolidSamplesPreparetheinstrumentasoutlinedintheoperatorsinstructionmanual.Determinetheinstrumentblank.Loginaminimumof3Blankreplicates.PresstheAnalyzebuttonontheinstrumentscreen.Afterashortdelay,theloadingheadslide-blockwillopen.Placea502-344NickelBasketintotheloadinghead.Note:samplesusingautomationshouldbeplacedintheappropriateautoloaderpositionbeforestartingtheanalysissequence.Oncethesequencehasstarted,theautomaticanalysiswillstartandendautomatically.LECO氮分析儀標(biāo)樣上海禹重實(shí)業(yè)有限公司致力于提供分析儀,期待您的光臨!

松江分析儀,分析儀

SummaryThedeterminationoftheamountofoxygen,nitrogen,andhydrogeniniron,steel,nickel-,andcobalt-basealloysrepresentssomeofthemostimportantqualitymetricsforthesematerials.Oxygenisusedtocreatesteelfrompigironbyremovingexcesscarbon.Oxygencontentmustbecontrolledtolimittheamountofcarbonmonoxidethatcanbeformedduringsolidificationwhichmaycauseexcessiveporosity.Nitrogenisconsideredbothanimpurityaswellasanimportantalloyingagent.Itcanbepresentasanitrideorinterstitiallyinitsgaseousform.Increasednitrogencontentisknowntoincreaseyieldandtensilestrength,thusdecreasingductilityandformability.Excessivelevelsmayevolveduringsolidificationthusincreasingporosity.Highhydrogencontentistheprimarycauseofembrittlement,blisteringandflakingduetoitshighmobilitythroughthelatticeandprovidesnopotentialalloyingbenefits.TheONH836utilizesahigh-powerelectrodefurnacetoquicklyandefficientlyreleasethetargetgasesfromwithinthesample,whichallowsforaveryrapidsimultaneousdeterminationofoxygen,nitrogen,andhydrogen.

Procedure–Powder/ChipSamplesDeterminetheinstrumentblank.PresstheAnalyzebuttonontheinstrumentscreenagain,theloadingheadslide-blockwillcloseandthelowerelectrodewillopen.Cleantheupperandlowerelectrodeeithermanuallyorremovethecrucibleandpresstheanalyzebuttontocleanwithanautomaticcleanerifapplicable.Addapproximately0.05gof***GraphitePowdertoa782-720GraphiteCrucible.Firmlyplacethecrucibleonthelowerelectrodetiporappropriateautoloaderposition.分析儀,就選上海禹重實(shí)業(yè)有限公司,用戶的信賴之選,有想法可以來我司咨詢!

松江分析儀,分析儀

氧氮?dú)浞治鰞x的常見故障及解決方法:氧和氮空白值超過20。這是由于氣流小,不能將爐子中的空氣驅(qū)趕出去??烧{(diào)節(jié)氣體流量,調(diào)節(jié)載氣壓力在0.2~0.4MPa。接通儀器載氣,放一個(gè)石墨坩堝在下電極上,打開主電源開關(guān),點(diǎn)擊軟件上的關(guān)爐按鈕,關(guān)閉爐子并等待10s。調(diào)節(jié)流量調(diào)節(jié)器,直到流量計(jì)a顯示為30L/h,打開爐子。調(diào)節(jié)調(diào)節(jié)器直到流量計(jì)b顯示為50L/h,再次關(guān)閉爐子。如果以上設(shè)置不穩(wěn)定,則增加流量至100L/h,反復(fù)調(diào)節(jié)直至儀器穩(wěn)定。分析儀,就選上海禹重實(shí)業(yè)有限公司,讓您滿意,歡迎您的來電!松江分析儀

分析儀,就選上海禹重實(shí)業(yè)有限公司,讓您滿意,有想法可以來我司咨詢!松江分析儀

SamplePreparationSamplingandsamplepreparationofrefractorymetalssuchastitaniumandzirconiumissomewhatdifferentfromthatofsteel.Unlikesteelsamples,hydrogenisnotasmobileinthisgroupofmaterials;therefore,storageinliquidnitrogenordryiceisnotrequired.However,itisimportanttokeepthesamplecoolwhencuttingorsectioning.Samplepreparationforoxygenandnitrogendeterminationhasbeendifferentfromthatofhydrogendetermination.松江分析儀

上海禹重實(shí)業(yè)有限公司是我國光譜儀,色譜儀,質(zhì)譜儀,樣品前處理專業(yè)化較早的有限責(zé)任公司(自然)之一,公司位于上海市閔行區(qū)春申路2525號(hào)3幢339室,成立于2013-10-22,迄今已經(jīng)成長為儀器儀表行業(yè)內(nèi)同類型企業(yè)的佼佼者。公司承擔(dān)并建設(shè)完成儀器儀表多項(xiàng)重點(diǎn)項(xiàng)目,取得了明顯的社會(huì)和經(jīng)濟(jì)效益。產(chǎn)品已銷往多個(gè)國家和地區(qū),被國內(nèi)外眾多企業(yè)和客戶所認(rèn)可。